Sub pages

Specification

Handyscope HS1 - Specification

Handyscope HS1

The Handyscope HS1 is no longer manufactured and sold.

Acquisition system
Number of input channels2 analog, BNC
Resolution12 bits = 0.025%
Accuracy0.25% ± 1 LSB
Ranges500 mVolt .. 20 Volt full scale
CouplingAC/DC, manual with switches on the front
Impedance1 MOhm / 20 pF
Protection1:1+ probe : 200 volt AC 50 Hz
1:10 probe : 600 volt AC 50 Hz
Bandwidth (-3dB)DC to 50 kHz maximum
AC coupling cut off frequency (-3dB)±1.5 Hz
Maximum sampling rate80 kHz on one channel, 40 kHz on two channels
Sampling sourcesoftware, ±5%
Memory-
Trigger system
Systemsoftware, 2 levels
SourceCH1, CH2, keyboard
Trigger modesrising slope, falling slope
Level adjustment0 - 100% of full scale
Hysteresis adjustment0 - 100% of full scale
Resolution0.39%, 8 bits
Pre trigger-
Post trigger-
Interface
InterfaceLPT
Power requirements
Power from LPT port120 mA max (0.6 Watt max)
Physical
Instrument height40 mm (1.5 inch)
Instrument length145 mm (5.7 inch)
Instrument width85 mm (3.3 inch)
Cord length1.7 meter (67 inch)
Weight350 gram (12.3 ounce)
Accessories2 oscilloscope probes 1:1 - 1:10 switchable