Sub pages

Specification

Handyscope HS2 - Specification

Handyscope HS2

The Handyscope HS2 is no longer manufactured and sold.

Acquisition system
Analog input channels2, BNC
Resolution12 bits = 0.025%
Accuracy0.2% ± 1 LSB
Ranges100 mVolt .. 80 Volt full scale
CouplingAC/DC
Impedance1 MOhm / 20 pF
Protection200 V (DC + AC peak) on 1:1 probe
600 V (DC + AC peak) on 1:10 probe
Bandwidth (-3dB)DC to 100 kHz maximum
AC coupling cut off frequency (-3dB)± 1.5 Hz
Maximum sampling rate200 kHz, 5 µsec
Sampling sourceinternal ±0.01%
Memory32760 samples/channel
Trigger system
Systemdigital, 2 levels
SourceCH1, CH2, keyboard
Trigger modesrising slope, falling slope, inside window, outside window
Level adjustment0 - 100% of full scale
Hysteresis adjustment0 - 100% of full scale
Resolution0.39% (8 bits)
Pre trigger0 - 32768 samples (0 - 100%, one sample resolution)
Post trigger0 - 32768 samples (0 - 100%, one sample resolution)
Interface
InterfaceLPT
Power Requirements
Power from LPT port200 mA max (1 Watt max)
Physical
Instrument height40 mm (1.5 inch)
Instrument length145 mm (5.4 inch)
Instrument width85 mm (3.3 inch)
Cord length1.8 meter (70 inch)
Weight500 gram (18 ounce)
Accessories2 oscilloscope probes 1:1 - 1:10 switchable
Options
Optionally available400 kHz maximum sampling frequency
Optionally availableDigital external trigger input