Sub pages

Specification

Handyscope HS4 - Specification

Handyscope HS4
Acquisition system
Number of input channels4 analog, BNC
Resolution12 bit @ 50 MS/s
14 bit @ 3.125 MS/s
16 bit @ 195.3125 kS/s
Accuracy0.2% ± 1 LSB
Ranges200 mVolt .. 80 Volt full scale
CouplingAC/DC
Impedance1 MOhm / 30 pF
Protection (in all ranges)±200 Volt (DC + AC peak < 10 kHz)
Bandwidth (-3dB)DC to 50 MHz maximum
AC coupling cut off frequency (-3dB)1 Hz with 1x probe
Maximum sampling rate50 MS/s, 20 nsec (model HS4-50)
25 MS/s, 40 nsec (model HS4-25)
10 MS/s, 100 nsec (model HS4-10)
5 MS/s, 200 nsec (model HS4-5)
Sampling sourceinternal, external
Sampling source internalquartz
Accuracy±0.01%
Stability±100 ppm over -40°C to +85°C
Memory0 - 131071 samples each channel
Trigger system
Systemdigital, 2 levels
SourceCH1, CH2, CH3, CH4, AND, OR, digital external
Trigger modesrising slope, falling slope, inside window, outside window, peak
Level adjustment0 - 100% of full scale
Hysteresis adjustment0 - 100% of full scale
Resolution0.025% (12 bits)
Digital trigger input level0 - 3.3 volt (5 volt max)
Pre trigger0 - 131071 samples (0 - 100%, one sample resolution)
Post trigger0 - 131071 samples (0 - 100%, one sample resolution)
Interface
InterfaceUSB 2.0 High Speed (480 Mbit/s); (USB 1.1 Full Speed (12 Mbit/sec) compatible)
Power Requirements
Power from USB port500 mA max (2.5 Watt max)
Power via external power input / extension connector1500 mA max (7.5 Watt max)
Minimum voltage4.5 Volt DC
Maximum voltage12 Volt DC
Physical
Instrument height25 mm (1 inch)
Instrument length170 mm (6.7 inch)
Instrument width140 mm (5.2 inch)
Cord length1.8 meter (70 inch)
Weight480 gram (17 ounce)
I/O connectors
Channel 1...4BNC
USBfixed cable with USB 2.0 and USB 1.1 type A
Extension connectorSub-D 25 pins female
System Requirements
PC I/O connectionUSB 2.0 High Speed (480 Mbit/s); (USB 1.1 Full Speed (12 Mbit/sec) compatible)
Operating SystemWindows 98/ME/2000/XP/Vista/7
Operating Environment
Ambient temperature0 to 55 °C
Relative humidity10 to 90% non condensing
Storage Environment
Ambient temperature-20 to 70 °C
Relative humidity5 to 95% non condensing
Certifications and Compliances
CE mark complianceyes
RoHSyes
Package
InstrumentHandyscope HS4
Probes4 oscilloscope probes 1:1 - 1:10 switchable
Accessoriesexternal power cable for PS/2 mouse or keyboard connector
SoftwareWindows 98/2000/ME/XP/Vista/7 on CD
DriversWindows 98/2000/ME/XP/Vista/7 on CD
Manualsinstrument manual and software user's manuals
color printed on paper and digital on CD