Accessories - Differential Probe - SI-9002

 
Description Measuring a power outlet

 

Description


The Differential Probe - SI-9002 extends the measurement power of oscilloscopes to power semiconductor circuits. This Differential Probe provides any general oscilloscope with high voltage differential input.

SI-9002 Image

Using a built-in differential amplifier, it scales and converts the high voltage differential input to a low voltage single ended BNC output. The differential input is useful in measurements of power semiconductor circuits, as no reference to ground is required. Both the positive and negative sides of the balanced input possess high voltage differential input up to 700 volt, and its sensitivity can extend down to 100 mvolt. With bandwidth and high response speed, it is adequate for accurate dv/dt measurement in the fastest switching circuits of thyristors, bipolar transistors, or power MOSFETS.

The Differential Probe - SI-9002 is small and light, so it can be used in a very convenient way; just plug it into the input of, for example, the Handyscope HS3. The wide bandwidth and the high accuracy of this Differential Probe cater for a wide range of applications.


 

Specification

 
Differential Probe 1:20-1:200 - SI-9002 - Specification
Bandwidth DC to 25MHz (-3dB)
Attenuation Ratio 1:20 or 1:200
Accuracy ±2%
Rise Time 14 ns
Input Impedance 4 Mohm
5.5 pF each side ground
Input Voltage
  Differential Range*
    1:20 ±140 V (DC+Peak AC) or 140 Vrms
    1:200 ±1400 V (DC+Peak AC) or 1000 Vrms
  Common Mode Range ±1400 V (DC+Peak AC) or 1000 Vrms
  Absolute Max. Voltage * ±1400 V (DC+Peak AC) or 1000 Vrms CAT III
Output
  Swing (into 2k ohm load) +/-7 V
  Offset (typical) <+/-5 mV
  Noise (typical) 0.7 mVrms
  Source Impedance (typical) 1 ohm @ 1 kHz, 8 ohm @ 1 MHz
CMRR (typical) -80 dB @ 50 Hz, -60 dB @ 20 kHz
Ambient
  Operating Temperature -10 to 40 °C (14 to 104 F)
  Storage Temperature -30 to 70 °C (-22 to 158 F)
  Operating Humidity 25 to 85% RH
  Storage Humidity 25 to 85% RH
Power Requirements Standard 4xAA cells
or regualted 9 VDC / 40mA mains adaptor**
Length of BNC Cable 95 cm (37.4 inch)
Length of Input Leads 45 cm (17.7 inch)
Weight (probe and PVC jacket) 400 gram (14.1 ounce)
Dimensions
  Height 21 mm (0.8 inch)
  Length 170 mm (6.7 inch)
  Width 63 mm (2.5 inch)
 
* Voltage limit is the lesser of the DC+Peak AC and RMS values.
** a. The supplied voltage must be less than 12V and greater than 4.4V, otherwise the probe could be damaged or can’t be operated properly.
b. polarity is “+” inside and “–” outside. For wrong polarity, built-in circuit protects the probe, no danger or damage will occur.
c. When the voltage of the cells become too low, the power indicator on the panel will flicker.
print