Home | Products | Software | Download | News | Search | F.A.Q. |        

Handyprobe HP2 - Hardware Specification

 
Description Hardware Specification Download

 

Hardware Specification


Acquisition system
Analog input channels 1, probe tip
    Resolution 8 bit = 0.39%
    Accuracy 1% ± 1 LSB
    Ranges 0.5 Volt .. 400 Volt full scale
    Coupling AC/DC
    Impedance 1 MOhm / 30 pF
    Protection 500 Volt
    Bandwidth (-3dB) DC to 2 MHz maximum
    AC coupling cut off frequency (-3 dB)± 1.5 Hz
Maximum sampling rate 20 MS/s,   50 nsec (model HP2-20MHz)
10 MS/s, 100 nsec (model HP2-10MHz)
5 MS/s, 200 nsec (model HP2-  5MHz)
2 MS/s, 500 nsec (model HP2-  2MHz)
1 MS/s,     1 µsec (model HP2-  1MHz)
Sampling source internal
    Accuracy 0.01%
    Stability ±100 ppm over -40°C to +85°C
Memory 65536 samples
 
Trigger system
System digital, 2 levels
Source channel, keyboard
Trigger modes rising slope, falling slope, inside window, outside window
Level adjustment 0 - 100% of full scale
Hysteresis adjustment 0 - 100% of full scale
Resolution 0.39% (8 bits)
Pre trigger 0 - 65535 samples (0 - 100%)
Post trigger 0 - 65535 samples (0 - 100%)
 
Interface
Interface LPT
 
Power Requirements
Power from LPT port 200 mA max (1 Watt max)
 
Physical
Instrument height 22 mm (0.8 inch)
Instrument length 125 mm (4.9 inch)
Instrument width 43 mm (1.7 inch)
Cord length 1.8 meter (70 inch)
Weight 260 gram (9.2 ounce)
 
I/O connectors
Channel 1 in probe tip
Channel 1 ground 13 cm wire with isolated crocodile clip
parallel port fixed cable with 25 pin male Sub-D connector
 
System requirements
PC I/O connection parallel port
Operating system Windows 95/98/NT4/2000/ME/XP
 
Certifications and compliences
CE mark compliance Yes
 
Package
Instrument Handyprobe HP2
Software Windows 95/98/NT4/2000/ME/XP on CD
Manuals instrument manual and software user's manuals
color printed on paper and digital on CD
 
Options
Optionally available banana sockets instead of probe tip and ground wire
 
print