| Acquisition system |
| Number of input channels | 2 analog, BNC |
| Resolution | 12 bits = 0.025% |
| Accuracy | 0.25% ± 1 LSB |
| Ranges | 500 mVolt .. 20 Volt full scale |
| Coupling | AC/DC, manual with switches on the front |
| Impedance | 1 MOhm / 20 pF |
| Protection | 1:1 probe : 200 volt AC 50 Hz 1:10 probe : 600 volt AC 50 Hz |
| Bandwidth (-3dB) | DC to 50 kHz maximum |
| AC coupling cut off frequency (-3dB) | ±1.5 Hz |
| Maximum sampling rate | 80 kHz on one channel, 40 kHz on two channels |
| Sampling source | software, ±5% |
| Memory | - |
| |
| Trigger system |
| System | software, 2 levels |
| Source | CH1, CH2, keyboard |
| Trigger modes | rising slope, falling slope |
| Level adjustment | 0 - 100% of full scale |
| Hysteresis adjustment | 0 - 100% of full scale |
| Resolution | 0.39%, 8 bits |
| Pre trigger | - |
| Post trigger | - |
| |
| Interface |
| Interface | LPT |
| |
| Power requirements |
| Power from LPT port | 120 mA max (0.6 Watt max) |
| |
| Physical |
| Instrument height | 40 mm (1.5 inch) |
| Instrument length | 145 mm (5.7 inch) |
| Instrument width | 85 mm (3.3 inch) |
| Cord length | 1.7 meter (67 inch) |
| Weight | 350 gram (12.3 ounce) |
| Accessories | 2 oscilloscope probes 1:1 - 1:10 switchable |
| |

|