| Acquisition system |
| Number of input channels | 2 analog, BNC |
| Resolution | 8 bit = 0.39% |
| Accuracy | 2% ± 4 LSB |
| Ranges | 500 mVolt .. 20 Volt full scale, each channel |
| Coupling | AC/GND/DC, manualy selected with switch on bracket |
| Impedance | 1 MOhm / 25 pF |
| Protection | 200 Volt AC 50 Hz (1:1 probe) 600 Volt AC 50 Hz (1:10 probe) |
| Bandwidth (-3dB) | DC to 100 kHz maximum |
| AC coupling cut off frequency (-3dB) | ± 1.5 Hz |
| Maximum sampling rate | 200 kHz, 5 µsec |
| Sampling source | software, ±5% |
| Memory | - |
| |
| Arbitrary Waveform Generator (independent from acquisition system) |
| Number of output channels | 1 analog, BNC |
| Resolution | 8 bit @ 200 kHz |
| Accuracy | 2% ± 4 LSB |
| Coupling | DC |
| Amplitude | ±1.25 Volt full scale and ±2.50 Volt full scale |
| Impedance | 50 Ohm |
| Bandwidth | DC to 200 kHz |
| Sampling rate | 200 kHz |
| Sampling source | software, ±5% |
| Memory | - |
| |
| Trigger system |
| System | software, 2 levels |
| Source | CH1, CH2, keyboard |
| Trigger modes | rising slope, falling slope |
| Level adjustment | 0 - 100% of full scale |
| Hysteresis adjustment | 0 - 100% of full scale |
| Resolution | 0.39% (8 bits) |
| Pre trigger | - |
| Post trigger | - |
| |
| Interface |
| Interface | 8 bit ISA slot |
| |
| Power Requirements |
| Power from ISA slot | 1.7 Watt max +5V 100mA +12V 50mA -12V 50mA |
| |
| Physical |
| Instrument height | 100mm (3.9 inch) |
| Instrument length | 125mm (4.9 inch) |
| Instrument width | 25mm (1 inch) |
| Weight | 110 gram (3.9 ounce) |
| |

|