Home | Products | Software | Download | News | Search | F.A.Q. |        

Handyscope HS3 - Hardware Specification

 
Description Hardware Specification Download

 

Hardware Specification


Acquisition system
Number of input channels 2 analog, BNC
    Resolution 8 bit ≤ 100 MS/s
12 bit ≤ 50 MS/s
14 bit ≤ 3.125 MS/s
16 bit ≤ 195 kS/s
    Accuracy 0.2% ± 1 LSB
    Ranges 200 mVolt .. 80 Volt full scale
    Coupling AC/DC
    Impedance 1 MOhm / 30 pF
    Protection (in all ranges) 200 Volt (DC + AC peak < 10 kHz)
    Bandwidth (-3dB) DC to 50 MHz maximum
    AC coupling cut off frequency (-3dB)1.5 Hz
Maximum sampling rate 100 MS/s,   10 nsec (model HS3-100)
50 MS/s,   20 nsec (model HS3-50  )
25 MS/s,   40 nsec (model HS3-25  )
10 MS/s, 100 nsec (model HS3-10  )
5 MS/s, 200 nsec (model HS3-5    )
Sampling source internal, external
Sampling source internal quartz
    Accuracy ±0.01%
    Stability ±100 ppm over -40°C to +85°C
Memory 131072 samples each channel
 
Arbitrary Waveform Generator (independent from acquisition system)
Number of output channels 1 analog, BNC
    Resolution 14 bit @ 50 MS/s
    Amplitude -12 Volt .. 12 Volt
    Accuracy 0.4%
    Amplitude step 0 - ±0.1 V range, 8192 steps
±0.1 - ±0.9 V range, 8192 steps
±0.9 - ±12 V range, 8192 steps
    Coupling DC
    Impedance 50 Ohm
    Bandwidth DC to 2MHz
    DC level 0 - ±12 V in 8192 steps
    Noise level 0 - ±0.1 V range : 900 µVolt RMS
±0.1 - ±0.9 V range : 1.3 mVolt RMS
±0.9 - ±12 V range : 1.5 mVolt RMS
Sampling rate 50 MHz
Sampling source internal
    Accuracy ±0.01%
    Stability ±100 ppm over -40°C to +85°C
Memory 1024 points DDS mode
128K points linear mode
Waveforms sine, triangle, square, DC, noise and user defined
Symmetry 1 - 99%, 1% steps
 
Trigger system
System digital, 2 levels
Source CH1, CH2, AND, OR, digital external
Trigger modes rising slope, falling slope, inside window, outside window, peak
Level adjustment 0 - 100% of full scale
Hysteresis adjustment 0 - 100% of full scale
Resolution 0.025% (12 bits)
Digital trigger input level 0 - 3.3 volt (5 volt max)
Pre trigger 0 - 131071 samples (0 - 100%, one sample resolution)
Post trigger 0 - 131071 samples (0 - 100%, one sample resolution)
 
Interface
Interface USB 2.0 High Speed (480 Mbit/s); (USB 1.1 Full Speed (12 Mbit/sec) compatible)
 
Power Requirements
Power from USB port 500 mA max (2.5 Watt max)
Power via external power input / extension connector1500 mA max (7.5 Watt max)
    Minimum voltage 4.5 Volt DC
    Maximum voltage (SN# <12941) 6 Volt DC
    Maximum voltage (SN# >12941) 12 Volt DC
 
Physical
Instrument height 25 mm (1 inch)
Instrument length 170 mm (6.7 inch)
Instrument width 140 mm (5.2 inch)
Cord length 1.8 meter (70 inch)
Weight 480 gram (17 ounce)
 
I/O connectors
Channel 1 ... 2 BNC
Generator BNC
USB fixed cable with USB 2.0 and USB 1.1 type A connector
Extension connector Sub-D 25 pins female
 
System requirements
PC I/O connection USB 2.0 High Speed (480 Mbit/s); (USB 1.1 Full Speed (12 Mbit/sec) compatible)
Operating system Windows 98/ME/2000/XP/Vista/7
 
Operating environment
Ambient temperature 0 to 55°C
Relative humidity 10 to 90%, non condensing
 
Storage environment
Ambient temperature -20 to 70°C
Relative humidity 5 to 95%, non condensing
 
Certifications and Compliances
CE mark compliance yes
RoHS yes
 
Package
Instrument Handyscope HS3
Probes 2 oscilloscope probes 1:1 - 1:10 switchable
Accessories external power cable for PS/2 mouse or keyboard connector
Software for Windows 98/2000/ME/XP/Vista/7 on CD
Drivers for Windows 98/2000/ME/XP/Vista/7 on CD
Manuals instrument manual and software user's manuals
color printed and digital on CD
 
Options
Optionally available 10 MHz generator output
Optionally available Single channel Handyscope HS3 without generator, but with 512K samples memory
 
print