| Acquisition system |
| Number of input channels | 2 analog |
| Ch1, Ch2 | BNC |
| Maximum sampling rate | 1 GS/s (one channel) 500 MS/s (two channels) |
| Sampling clock source | internal, external |
| Internal | quartz |
| Accuracy | ±0.01% |
| Stability | ±100 ppm over -40°C to +85°C |
| External | LVDS, on extension connector |
| Memory | 32 MiSamples per channel |
| |
| Inputs Ch1 and Ch2 |
| Measure mode | Voltage |
| Resolution | 8 bits |
| Amplitude accuracy | 1% |
| Offset accuracy | 1% |
| Ranges | 200 mV .. 80 V full scale in a 2 - 4 - 8 sequence |
| Coupling | AC/DC |
| Impedance | 1 MOhm // 20 pF |
| Maximum input voltage (in all ranges) | 200 V (DC + AC peak < 10 kHz) |
| Maximum input voltage with 1:10 probe (in all ranges) | 600 V (DC + AC peak < 10 kHz) |
| Bandwidth (-3dB) | 250 MHz |
| AC coupling cut off frequency (-3dB) | 1.5 Hz |
| Rise time | 1.5 ns |
| Overshoot | 1% |
| |
| Trigger system |
| System | digital, 2 levels |
| Source | CH1, CH2, AND, OR, Analog external, Digital external, AWG Start, AWG Stop, AWG New Period |
| Trigger modes | rising slope, falling slope, inside window, outside window |
| Level adjustment | 0 - 100% of full scale |
| Hysteresis adjustment | 0 - 100% of full scale |
| Resolution | 0.39% (8 bits) |
| Pre trigger | 0 - 32MiSamples (0 - 100%, one sample resolution) |
| Post trigger | 0 - 32MiSamples (0 - 100%, one sample resolution) |
| Analog external trigger | |
| Input | BNC |
| Range | 200 mV .. 80 V full scale |
| Coupling | AC/DC |
| Bandwidth | 20 MHz |
| Maximum input voltage (in all ranges) | 200 V (DC + AC peak < 10 kHz) |
| Digital external trigger | |
| Input | extension connector |
| Range | 0 .. 5 V (TTL) |
| Coupling | DC |
| |
| Arbitrary Waveform Generator (independent from acquisition system) |
| Output channel | 1 analog, BNC |
| Resolution | 14 bits @ 200 MHz |
| Amplitude | -12 .. +12 V |
| Amplitude step | 0 .. 0.12 V range: 12 bit resolution 0.12 .. 1.2 V range: 12 bit resolution 1.2 .. 12 V range: 12 bit resolution |
| Coupling | DC |
| Impedance | 50 Ohm |
| Bandwidth | DC to 20 MHz |
| DC Level | -12 .. +12 V, 12 bit resolution |
| Maximum sampling rate | 200 MHz |
| Sampling clock source | internal, 0.01% |
| Accuracy | ±0.01% |
| Stability | ±100 ppm over -40°C to +85°C |
| Memory | 32 MiSamples |
| Waveforms | Sine, Square, Triangle, DC, Noise, Arbitrary |
| Symmetry | 0 .. 100% |
| |
| Interface |
| Interface | USB 2.0 High Speed (480 Mbit/s) USB 1.1 Full Speed (12 Mbit/sec) compatible |
| |
| Power Requirements |
| Power adapter | External |
| Input | 110 .. 240 V AC, 50 .. 60 Hz 0.85 A Max., 50 VA .. 80 VA |
| Output | 5 V DC, 1.5 A 12 V DC, 1.5 A |
| Dimension | 33 x 58 x 102 mm / 1.3 x 2.3 x 4 " (h x w x l) |
| Consumption | 5 V DC, 1500 mA max 12 V DC, 1500 mA max |
| |
| Physical |
| Instrument height | 57 mm (2.3 inch) |
| Instrument length | 276 mm (10.9 inch) |
| Instrument width | 180 mm (7.1 inch) |
| Weight | 1250 gram (2.8 lbs) |
| USB cable length | 1.2 meter (47 inch) |
| Power cord length | 1.5 meter (59 inch) |
| |
| I/O connectors |
| Ch1, Ch2 | BNC |
| Generator out | BNC |
| Analog external trigger input | BNC |
| Power | 5 pin Mini-DIN socket |
| USB | type B socket |
| Extension connector | Sub-D 15 pins female |
| |
| System Requirements |
| PC I/O connection | USB 2.0 High Speed (480 Mbit/s) USB 1.1 Full Speed (12 Mbit/s) compatible |
| Operating System | Windows 98/ME/2000/XP/Vista/7 |
| |
| Operating Environment |
| Ambient temperature | 0 to 55 °C |
| Relative humidity | 10 to 90% non condensing |
| |
| Storage Environment |
| Ambient temperature | -20 to 70 °C |
| Relative humidity | 5 to 95% non condensing |
| |
| Certifications and Compliances |
| CE mark compliance | Yes |
| RoHS | Yes |
| |
| Package |
| Instrument | TiePieSCOPE HS805 |
| Probes | 2 x Probe 1:1-1:10 - HP-9250 |
| Software | Windows 98/2000/ME/XP/Vista/7 on CD |
| Drivers | Windows 98/2000/ME/XP/Vista/7 on CD |
| Manual | instrument manual and software user's manuals color printed on paper and digital on CD |
| |