| Acquisition system |
| Number of input channels | 2 analog, BNC |
| Resolution | 8 bit = 0.39% |
| Accuracy | 1% ± 1 LSB |
| Ranges | 100 mVolt .. 80 Volt full scale |
| Coupling | AC/DC |
| Impedance | 1 MOhm / 30 pF |
| Protection (in all ranges) | 200 Volt (DC + AC peak < 10 kHz) |
| Bandwidth (-3dB) | DC to 50 MHz |
| AC coupling cut off frequency (-3dB) | ±1.5 Hz |
| Maximum sampling rate | 100 MHz, 10 nsec on one channel 50 MHz, 20 nsec on two channels |
| Sampling source | internal ±0.01%, external |
| Memory | 65536 samples each channel |
| |
| Arbitrary Waveform Generator (independent from acquisition system) |
| Number of output channels | 1 analog, BNC |
| Resolution | 10 bit |
| Amplitude | -10 Volt .. 10 Volt |
| Amplitude step | 0 - 0.1 V range, 4096 steps 0.0 - 0.9 V range, 4096 steps 0.9 - 10 V range, 4096 steps |
| Coupling | DC |
| Impedance | 50 Ohm |
| Bandwidth | DC to 2MHz |
| DC level | 0 - 10 V in 4096 steps |
| Sample rate | 0 - 25 MHz |
| Sampling source | internal ±0.01% |
| Memory depth | 1024 points DDS mode 65535 points linear mode |
| Waveforms | sine, triangle, square, DC, noise and user defined |
| Symmetry | 1 - 99%, 1% steps |
| |
| Trigger system |
| System | hardware, 2 levels |
| Source | CH1, CH2, AND, OR, digital external |
| Trigger modes | rising slope, falling slope, inside window, outside window, TV line, TV frame |
| Level adjustment | 0 - 100% of full scale |
| Hysteresis adjustment | 0 - 100% of full scale |
| Resolution | 0.39% (8 bits) |
| Digital trigger input level | 0 - 5 V max |
| Pre trigger | 0 - 65520 samples (0 - 100%) |
| Post trigger | 0 - 65520 samples (0 - 100%) |
| |
| Interface |
| Interface | 8 bit ISA slot |
| |
| Power Requirements |
| Power from ISA slot | 3.5 Watt max |
| |
| Physical |
| Instrument height | 127 mm (4.9 inch) |
| Instrument length | 175 mm (6.8 inch) |
| Instrument width | 22 mm (0.9 inch) |
| Weight | 150 gram (5.3 ounce) |
| Accessories | 2 oscilloscope probes 1:1 - 1:10 switchable bracket with sub-D connector |
| |