TiePie engineering measuring instruments can work together with DASYLab. This extends the possibilities of the TiePie engineering measuring instruments even further.

Preparation

Files

The DASYLab driver package contains the following files:

Name    Description
TPDasy.dll    the driver
TPDasy.ini    INI file for the driver
Install.pdf    document with installation instructions

Installation

Copy the following files to the directory where DASYLab is installed:

  • TPDasy.dll
  • TPDasy.ini
  • The DLL for your instrument and the HEX files that come with it

Modify DASYLab.ini

Locate the file DASYLab.ini. In DASYLab version 9 and lower, this file is located in the application directory where DASYLab is installed.

Starting with DASYLab version 10, this file is located in a different folder on the computer, to meet the Windows Vista requirements:

Users\Public\Public Documents\DASYLab\10.0.0\ENG

On Windows XP, that folder is located at

Documents and Settings\All Users\Documents\DASYLab\10.0.0\ENG

In the file DASYLab.ini, change or add the following lines:

[Hardware]
Driver=TPDasy.dll
TPDasy.dll=TiePie instrument
 
[Extend]
DLL=TPDasy.dll

Modify TPDasy.ini

If you use a different instrument than a Handyscope HS3, you will have to change the DLL name in TPDasy.ini to the name of the DLL for your instrument. By default the name is set to:

DLLName=hs3.dll

Using the instrument in DASYLab

DASYLab version 10

Start DASYLab. Then go to Measurement -> Driver selection... and select the TiePie instrument. You can now use the instrument.

You can find it under Modules -> Input/Output -> Driver -> Analog input. Drag this object onto the worksheet. When you double-click it, a dialog will pop up in which you can enable different channels.

Time base settings can be changed in Measurement -> Time Bases -> All Settings....

To change other instrument settings, go to Measurement -> Driver....

DASYLab version 9 and older

Start DASYLab. Then go to Experiment -> Select driver and select the TiePie instrument. You can now use the instrument.

You can find it under Modules -> Input/Output -> Driver -> Analog input. Drag this object onto the worksheet. When you double-click it, a dialog will pop up in which you can enable different channels.

Time base settings can be changed in Experiment -> Time Base Setup -> All Settings....

To change other instrument settings, go to Experiment -> Driver....